Let’s work through this timeout issue— I’ve run into the exact same problem with Agilent power supplies before, so here are the most likely fixes to try:
1. Double-Check You’re Connecting to the Right Device
Your code uses res[-1] to pick the last resource in the list, but that might not be your Agilent power supply. First, look at the output of print(res)— Agilent USB devices usually look like USB0::0x0957::0xXXXX::MYXXXXXXX::INSTR, while GPIB devices will start with GPIB0::.
Instead of relying on res[-1], manually specify the correct resource string to eliminate this variable:
inst = rm.open_resource('USB0::0x0957::0x0607::MY12345678::INSTR') # Replace with your actual address
2. Adjust Termination Characters
Agilent instruments often use \r\n (carriage return + newline) as the termination character instead of just \n. Your current setup uses \n, which might be causing the instrument to not recognize the command, or PyVISA to never detect the end of the response. Try updating these lines:
inst.write_termination = '\r\n' inst.read_termination = '\r\n'
3. Split Query into Write + Read to Isolate the Issue
The query() method combines writing the command and reading the response, which can make it hard to tell where the problem is. Split it up to test each step separately:
inst.write('*IDN?') import time time.sleep(1) # Give the instrument a second to respond print(inst.read())
If the write succeeds but the read times out, the instrument either didn’t receive the command, isn’t configured to respond, or you’re on the wrong address.
4. Verify Hardware and Instrument Power
- If using USB: Unplug and re-plug the USB cable, make sure the power supply is turned on, and check that your PC recognizes the device (no yellow exclamation marks in Device Manager).
- If using GPIB: Confirm the GPIB card is installed correctly, and that the instrument’s GPIB address matches what’s in your resource string.
5. Update PyVISA and Its Backend
Outdated versions of PyVISA or its backend (like pyvisa-py or NI-VISA) can cause compatibility issues with Agilent gear. Update them with:
pip install --upgrade pyvisa pyvisa-py
If you’re using NI-VISA instead of pyvisa-py, make sure your NI-VISA driver is up to date (check the NI software manager for updates).
6. Check the Instrument’s Remote Control Settings
Some Agilent power supplies require you to enable remote control from the front panel. Look for a menu option like "Remote" or "USB/GPIB Remote" and make sure it’s turned on— without this, the instrument won’t respond to commands sent over the interface.
If none of these work, try testing the communication with NI MAX (if you have NI-VISA installed) or Agilent’s IO Library Suite. If those tools also time out when sending *IDN?, the problem is likely with your hardware or instrument configuration, not PyVISA.